CSA 221421
Ex ia IIC T3/T4 Ga; Ex ia I Ma
AEx ia IIC T3/T4 Ga
CLI GPS A,B,C,D
CLII, GPS E,F,G CLIII
CLI, Zone 0
Operating Temp Code: T3
Ambient Temp Range = -40°C TO +121°C
Operating Temp Code: T4
Ambient Temp Range = -40°C TO +80°C
Control Drawing INS10012
Ex ia IIC T3 -40°C < Ta < +121°C
Ex ia IIC T4 -40°C < Ta < +80°C
1600 ft (488 m) Maximum Allowable
Integral Cable Length
Approved Cabling: CB112, CB191*, CB298 (*Approved for Use in T4 Applications Only)
Requires One IS111-1B Intrinsic Safety Barrier Per Sensor
IECEx SIR 15.0060X
Ex ia IIC T3-T4 Ga
Ex ib IIIC T135°C … T147°C Db
Ex ia I Ma
T3: Ta @ -40°C to +121°C
T4: Ta @ -40°C to +80°C
Mining Group I: Ta @-40°C to +121°C
T147°C: Ta @ -40°C to +121°C
T135°C: Ta @ -40°C to +80°C
Install per drawing INS10210
Refer to product manual MNX10134
Sira 15ATEX2152X
Ex ia IIC T3 -T4 Ga
T3: Ta @ -40°C to +121°C
T4: Ta @ -40°C to +80°C
Ex ib IIIC T135°C … T147°C Db
T147°C: Ta @ -40°C to +121°C
T135°C: Ta @ -40°C to +80°C
Ex ia I Ma
Ta @-40°C to +121°C
Install per drawing INS10211
Refer to product manual MNX10135
Connector Exit Entity Parameters
Ui= 28 VDC Ii = 120mA
Ci = 1.5nF Li = 0uH Pi = 1W
Integral Cable Entity Parameters
Ui= 28 VDC Ii = 120mA
Ci = 82nF Li = 336uH Pi = 1W